This is an open-access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in the Interactive Journal of Medical Research, is properly cited. The complete bibliographic information, a link to the original publication on http://www.i-jmr.org/, as well as this copyright and license information must be included.
New technology for clinical data collection is rapidly evolving and may be useful for both researchers and clinicians; however, this new technology has not been tested for accuracy, reliability, or validity.
This study aims to test the accuracy of visual analog scale (VAS) for pain on a newly designed application on the iPad (iPadVAS) and measure the reliability and validity of iPadVAS compared to a paper copy (paperVAS).
Accuracy was determined by physically measuring an iPad scale on screen and comparing it to the results from the program, with a researcher collecting 101 data points. A total of 22 healthy community dwelling older adults were then recruited to test reliability and validity. Each participant completed 8 VAS (4 using each tool) in a randomized order. Reliability was measured using interclass correlation coefficient (ICC) and validity measured using Bland-Altman graphs and correlations.
Of the measurements for accuracy, 64 results were identical, 2 results were manually measured as being 1 mm higher than the program, and 35 as 1 mm lower. Reliability for the iPadVAS was excellent with individual ICC 0.90 (95% CI 0.82-0.95) and averaged ICC 0.97 (95% CI 0.95-1.0) observed. Linear regression demonstrated a strong relationship with a small negative bias towards the iPad (−2.6, SD 5.0) with limits of agreement from −12.4 to 7.1.
The iPadVAS provides a convenient, user-friendly, and efficient way of collecting data from participants in measuring their current pain levels. It has potential use in documentation management and may encourage participatory healthcare.
Australia New Zealand Clinical Trials Registry (ANZCTR): 367297; https://www.anzctr.org.au/Trial/Registration/TrialReview.aspx?id=367297&isReview=true (Archived by Webcite at http://www.webcitation.org/6d9xYoUbD).
The health care sector is poised at the cusp of a transformation from being reactive to disease and injury toward proactive prevention, where the ultimate goal is to maximize individual health rather than treat disease. Ready access to medical information combined with ubiquitous sensing, quantified self, mobile computing, and social networking technologies empowers individuals to participate in their own health and well-being. According to Hood and colleagues [
The ability to measure pain objectively forms an important part of health care, both in chronic health monitoring and in acute settings, to determine changes in patient clinical presentation and the effectiveness of interventions aimed at alleviating pain. Visual analog scales (VAS) for collecting pain data in the traditional paper-based format have been shown to be accurate, valid, reliable, and reproducible [
Collecting the pain data electronically streamlines data measurement and management. Previously, electronic data collection using hand-held devices (eg, personal digital assistants or laptop computers) for VAS for pain has found values to be equivalent to paper-based tools; however, these electronic tools were costly [
Costs for new technology including hand-held tablets have decreased in recent years. These devices have the benefit of a user-friendly touch screen interface. With appropriate applications, data collected on a touch screen can be automatically measured and exported to a database for secure storage within the device and can easily be emailed to the researcher or clinician when access to the Internet is available. However, this new technology has not been tested for accuracy and reliability or compared to the paper-based gold standard for validity.
We assessed the accuracy of VAS for pain on an iPad (iPadVAS), measured and compared the reliability of iPadVAS to a paper copy (paperVAS), and validated the iPadVAS against paperVAS in a healthy community group.
A single researcher drew a line across the iPadVAS line with 1 finger 100 times using all parts of the scale. After each effort, the value was measured on the 100 mm line with a ruler that had 1 mm gradations marked on it; the number corresponding to the value of the mark (ie, a number between 0 and 100) was recorded in an Excel spreadsheet (Microsoft Corp). The researcher was blinded to the results generated by the application at the time. Measures by the researcher were then compared to the data produced by the algorithm in the iPad application.
The paperVAS was administered mounted on a clipboard and completed using a pen (0.7 mm tip width) on a line 100 mm in length and 0.75 mm high with no markings on the scale except
iPadVAS was administered on an iPad 2 and completed by the participant using their finger on the screen using an application developed by the research team [
Screen capture of a blank iPadVAS, a completed iPadVAS, and the CSV data output file.
This was a single center study conducted in healthy older adults (ages 65-85 years) at the Exercise Physiology Clinic at the University of Tasmania, Launceston, Australia. Participants who were enrolled for group Pilates classes were invited to take part in the study. The exclusion criterion was people who self-reported inadequate vision to complete the tasks. This study was approved by Human Medical Research Ethics Committee (Tasmania) Network (H0014062). The study is registered with ANZCTR (367297). Written informed consent was obtained from each participant (see
The study involved two tasks: completion of a paper-based and an electronic VAS for pain. Tests were administered in a randomized order in a single session lasting 30 minutes. Data were collected between November 2014 and April 2015. Each participant chose a token with eyes closed and the color of token selected determined the order (blue: iPadVAS first; red: paperVAS first) of the tasks. Each study participant was given instruction to draw a line through the line on the paper or iPad that corresponded to their current level of pain. They were provided with a demonstration of both tools. Each participant completed both tasks four times. After each effort, the results for the task on iPadVAS were saved and the screen reset so that the previous data were not available for comparison to the participant. Similarly, for the paperVAS, information from previous efforts was not available to the participants.
A change of 13 points in VAS for pain is considered as a clinically significant change [
All analyses were performed using Stata Intercooled software version 13 (StataCorp LP). The accuracy of the application algorithm to determine the value on the scale was analyzed by comparing data manually measured using a paired
Reliability was measured for both the iPadVAS and paperVAS using absolute agreement interclass correlation coefficient (ICC, 95% CI), and linear regression was used to determine the relationship between these two methods of data collection. Reliability was reported as excellent (ICC 0.90 and higher), good (ICC between 0.80 and 0.89), moderate (ICC between 0.70 and 0.79), or low (ICC less than 0.70) [
A researcher compared 101 data points by examining the difference between manual measurement and the calculated measurement produced by the iPad application program. Accuracy was high with 64 identical results, 2 manually measured results 1 mm higher, and 35 lower by 1 mm than the iPad program. Student
Linear regression showed high correlation of the scores between the two measurement techniques (
Correlation of VAS scores between manual and iPad application program.
A total of 22 community dwelling older adults (4 men, ages 56-86 years) were recruited to test reliability and validity. Grouped (iPad and paper) mean (SD) scores for pain values were 11.9 (10.6). Reliability for both tools was excellent (
There was a small negative bias (SD of bias) toward the iPad (−2.6 [5.0]) with limits of agreement between −12.42 and 7.14 (
Absolute agreement ICC for the two tools.
Variable |
paperVAS |
iPadVAS |
Individual | 0.96 (0.92-0.98) | 0.90 (0.82-0.95) |
Average | 0.99 (0.98-0.99) | 0.97 (0.95-1.00) |
Correlation of scores between paperVAS and iPadVAS measures of pain. Each data point is mean of 4 readings for each participant (data points 22).
Bland-Altman graph with difference and average of paperVAS and iPadVAS.
This is the first study to measure accuracy, reliability, and validity of an application on a touch screen iPad for VAS. Accuracy and reliability of the iPadVAS is excellent. Validity shows a small negative bias, but the value of this is not clinically relevant. The iPad application is accurate in that the program reads the same as a manual measurement with a ruler and has a correlation coefficient of .99.
A strength of using the iPad is that it prevents people scoring a line outside the VAS line, which consequently prevents invalid results from being recorded. When using a paper-based version of the VAS, some people indicate their pain levels by drawing less than 0 or more than 100 on the paper. These results can be either interpreted as 0 or 100, respectively, or considered an invalid result. In addition, the thickness of the line drawn by the participant is not affected by finger or stylus width. The high level of similarity of results between paperVAS and iPadVAS indicates that the iPadVAS is a clinically useful tool for collecting both individual and group data.
The magnitude of bias detected in our Bland-Altman analysis is not of clinically relevant amount. A minimum of 13 mm change is required over time to suggest if pain has increased or decreased [
This new tablet technology is superior to previous electronic data collection tools. The difficulty with personal digital assistants was that the full 100 mm standardized scale could not be used because of the small screen size [
The iPad data collection method has several strengths including portability with large data storage capacity, the ability to simply use Internet access to send data to the health care practitioner, and the potential to interface with other medical records. These features in combination with reduced costs demonstrate that this tool may have the potential to facilitate communication between clinicians and clients while enhancing participatory health care.
For clinicians and researchers, especially those involved in field work, the time, cost, and space savings of data storage are large compared to paper-based copies requiring manual measurement of values and transcription into databases or clinical notes. Hand-held electronic devices collecting questionnaire data show improved documentation completeness and fewer errors than paper-based counterparts [
Reported benefits for data management [
One benefit of using this technology may be the ability of the devices to provide individuals with a means to objectively monitor and record their pain status without requiring them to attend physical consultation. This is especially important for geographically isolated people and those with limited mobility. The ease of frequent monitoring without the need for recall between consultations may also facilitate regular remote monitoring of chronic health conditions. Similar to other Web-based resources [
A limitation of our study is that the data were collected in a sample of people from the community who were not in high levels of pain. Replication of this study in participants with moderate to high levels of pain would establish reliability in that population, although previous research indicates that the minimally clinically significant difference in pain scales such as VAS does not differ in populations with different severities of pain [
The iPadVAS provides a convenient, user-friendly, and efficient way of collecting data from participants in measuring their current pain levels. Its use in health care documentation management has the potential to encourage participatory health care. It is accurate, reliable, and valid in healthy older adults.